Note: Calibration should only be attempted if a 94dB calibrator is available. Attempting a calibration without a calibrator on the microphone will result in a complete decalibration of the instrument. When that occurs the factory must be contacted to provide the instrument’s original calibration file.
The instrument can be field-calibrated using a 94 dB calibrator with a standard 1/2” opening.
Field calibration is performed by the Instrument_Manager application. Follow these steps:
1. Remove the wind-screen
2. Place the instrument’s microphone in the calibrator
3. Start the calibrator
4. Press the Calibrate button on the application window and follow the on-screen instructions.
The calibration button can also be used to recall the factory calibration in case it is lost, or the instrument is decalibrated by mistake.